### TL;DR
A MEMS-based atomic force microscope (AFM) is used to image nanoscale features, including bacteria and silver nanoprisms, with a high-speed video system capturing the 9 kHz tapping motion in slow-motion via stroboscopic illumination.
## Atomic Force Microscope: A Compact MEMS-Based System
The AFM shown in this video is a small, easy-to-use device that generates sub-micron or even single-molecule images by tapping the sample surface. Unlike conventional AFMs that require optical alignment and a laser to detect cantilever deflection, this microscope is built entirely from MEMS (Micro-Electro-Mechanical Systems). The tip is a tiny chip that handles X, Y, and Z motion, plus an oscillator to drive the probe at around 9 kHz. The entire tip assembly is replaceable—if damaged, just swap the chip and the system is ready again. This design makes setup extremely simple: place the sample, raise the stage until the tip contacts the surface, and begin scanning. The software adjusts the tip height to maintain constant tapping oscillation amplitude, generating images based on nanoscale height differences.
The AFM is highly sensitive to airflow and vibration. A stationary garage with all windows and doors closed is sufficient, but vibration isolation is critical. The presenter uses a heavy granite slab supported on inner tubes to dampen vibrations. Even then, a light tap on the table creates disturbances that cause scanning lines to break and corrupt the image.
## High-Speed Video of the AFM Tip in Action
To see the tapping motion of the AFM tip, a Kronos stroboscopic high-speed camera system was used. The tip oscillates at about 9 kHz with an amplitude of a few microns. By pulsing the strobe light at a slightly different frequency (e.g., 8.5 kHz instead of 9 kHz), the tip motion appears slowed down or frozen. The video shows the tip moving, though the number of taps per scan is reduced by roughly a hundred times compared to real-time (actual tapping is 9,000 times per second, and a full scan takes minutes). Even so, capturing the tapping motion was a major achievement.
## Bacteria Imaging: Natto as a Sample
The goal was to image bacteria—ideally rod-shaped, large enough to be visually interesting, harmless, and easy to obtain. Natto (fermented soybeans) was chosen. The bacteria are on the soybean surface, and the sticky biofilm they produce makes them easy to isolate.
### Sample Preparation
1. **Growth medium**: Ingredients are mixed, boiled, and sterile-filtered through a 0.2 µm filter.
2. **Culturing**: A small piece of natto is added to the sterile medium, mixed, and left for a day or two. Bacteria detach from the soy and become free-floating.
3. **Centrifugation and washing**: The mixture is centrifuged to pellet the bacteria. The supernatant is discarded and replaced with clean water or very dilute saline. This washing step is repeated to remove salts and other dissolved solids that would create crystals and obscure AFM images.
4. **Final pellet**: Clean bacteria in pure water.
### Fixing Bacteria on a Surface
The challenge is to immobilize bacteria without clumping them. A gelatin coating on a silicon wafer works well. Steps:
- Break a silicon wafer into small pieces and glue them onto glass slides.
- Make the silicon surface hydrophilic: using a corona-discharge pen or air plasma (creates a reactive plasma that makes the surface water-friendly).
- Prepare a very dilute gelatin solution (plain gelatin from a grocery store), filter through a 0.2 µm filter to remove dust particles.
- Spin-coat the gelatin onto the silicon: a thin layer (~tens of nm) is formed. Color changes are visible as the layer thins.
- Add the cleaned bacterial suspension onto the gelatin-coated silicon. Wait 5–30 minutes. Gravity pulls bacteria down, and electrostatic attraction (opposite charges between gelatin and bacteria) binds them.
- Spin off excess liquid and wash repeatedly to remove any non-adherent particles.
- Dry the sample.
The resulting images show bacteria that appear rod-shaped with a "nose" protruding—almost like a hippopotamus head. The orientation is consistent, possibly due to the AFM tip's angle (not perfectly 90°) or the spin-coating direction. The MEMS AFM requires dry samples (water would damage the MEMS mechanism). The presenter considered high-humidity non-condensing environments but did not pursue it.
## Silver Nanoprisms: Imaging with AFM
Silver nanoprisms were created earlier by growing silver nanoparticles from a seed solution using bright LED illumination for several days. The nanoparticles absorb specific wavelengths and grow, changing the solution color: blue light yields yellow solution (blue removed), green yields purple/blue (green removed). The nanoprisms are thin (~10–20 nm) and have a diameter between 1 and 200 nm. They were fixed on gelatin-coated silicon using the same method. The images were somewhat ambiguous due to the small thickness, but the gelatin technique worked.
---
**Source:** [Atomic Force Microscope high-speed video, stainless etching, bacteria, and more](https://www.youtube.com/watch?v=DyIQkqBXhS0) – mhb